Abstract

This study evaluated the Crib-O-Gram as a screening test for hearing loss in NICU infants using Brain Stem Electric Response Audiometry (BERA) as the standard. Two Crib-O-Gram tests were administered to 280 babies within 48 hours of BERA testing. An additional 26 babies received BERA and one Crib-O-Gram test. Correlation coefficients for the two Crib-O-Gram scalar scores were calculated separately for a group of preterm (31 to 37 weeks) and a group of full term and older (38 to 54 weeks) infants, with values of 0.36 and 0.52, respectively. This indicates that the Crib-O-Gram has poor reliability for a screening test, particularly for preterm babies. Approximately one-third of babies with normal BERA thresholds failed Crib-O-Gram screening. The identification of hearing loss by Crib-O-Gram increased with the severity of the loss. Crib-O-Gram was able to identify moderately severe hearing losses in full term and older babies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.