Abstract

The prevalence of chronic renal diseases in Indonesia has an increasing annual trend, because it is frequently unrecognized and often co-exists with other disease. GFR and ERPF are parameters currently utilized to estimate renal function at routine renal scintigraphy 99m-Tc DTPA study. This study used 99m-Tc DTPA to measure GFR and ERPF. The purpose of this study was to find the correlation between ERPF and GFR, for ERPF analysis with Schlegel's method, and GFR analysis with Gate's method, as well as to find correction factor between both variables. Analysis of renal scintigraphy has been performed at Department of Nuclear Medicine Pertamina Center Hospital to thirty patient images acquired from 2014 to 2015 which were analyzed retrospectively data, using gamma camera dual head with counting method from renal scintigraphy 99m-Tc DTPA study. The calculation was executed by means of both display and manual calculation. Pearson's statistical analysis resulted on Positive Correlation for all data, with ERPF and GFR (display) showing Strongly Positive Correlation (r = 0.82; p- value < 0.05). Standard deviation was found to be 27.58 and 107.64 for GFR and ERPF (display), respectively. Our result indicated that the use of 99mTc-DTPA measure ERPF was not recommended.