Abstract

It has been established that while the inherent presence of bubbles increases heat generation due to scattering and absorption, inertial cavitation is responsible for elevated heating during high-intensity focused ultrasound (HIFU) application. The contribution of bubble-induced heating can be an important factor to consider, as it can be several times greater than the expected heat deposition from absorption of energy from the primary ultrasound field. The temperature and cavitation signal near the focus were measured for 5.5-s continuous-wave 1.1-MHz HIFU sonications in tissue mimicking phantoms. The measured temperature was corrected for heating predicted from the primary ultrasound absorption to isolate the temperature rise from the bubble activity. The temperature rise induced from cavitation correlates well with a measurement of the instantaneous "cavitation power" as indicated by the mean square voltage output of a 15-MHz passive cavitation detector. The results suggest that careful processing of the cavitation signals can serve as a proxy for measuring the heating contribution from inertial cavitation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.