Abstract

A high resolution electrical finite difference model of the human thorax based on a 43 slice MRI data set along with lead field theory was used to examine the contribution of the lungs to the total impedance for a typical mid-thoracic 2D EIT eight and sixteen electrode configuration. Regional analysis of the thoracic sources of impedance revealed that the maximum contribution of lungs to the total impedance was approximately 22% for the eight electrode array and 25% for the sixteen electrode array. Analysis of impedance distribution of the lungs using a mid-thoracic application showed that the contribution of impedance of each slice followed closely the volume of the lungs in the given slice. This suggests that the mid-thoracic application gives results reflecting the entire lung. The contributions of the lung impedance for the various electrode positions showed that the eight electrode configuration had a more smooth change between adjacent electrodes compared to the 16 electrode arrangement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.