Abstract

AbstractThe aim of this study is to build and test an amperometric biosensor to determine oxidized glutathione (GSSG). The biosensor is based on glutathione reductase (GR), which was immobilized onto the surface of reduced graphene oxide (rGO) modified with cobalt sulfide (CoS2). CoS2 nanostructures on the composite paper surface are active centers for immobilizing the GR enzyme. The conversion of GSSG into reduced glutathione was monitored with voltammetry in the presence of nicotinamide adenine dinucleotide phosphate reduced form (NADPH) on the GR‐CoS2/rGO immobilized electrode surface. The biosensor's cyclic volt response is linearly correlated with GSSG concentrations of 0.01–500 μM in a 0.1 M pH 7.4 phosphate‐buffer solution (PBS). The performance of the sensor was high, with a detection limit of 0.48 μM and a sensitivity of 0.6 mA μM cm−2. Four human blood samples were tested for GSSG with both the biosensor and Ellman's assay, which showed nearly identical results, confirming the biosensor's reliability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.