Abstract

Atrial fibrillation (AF) is the leading cause of stroke. The European Society of Cardiology (ESC) advises opportunistic AF screening among patients aged ≥ 65 years. Considering this, the aim herein, was compare the feasibility of two different systems of smartphone-based electrocardiogram (ECG) recordings to identify AF among those without a previous arrhythmia history. Prospective AF screening was conducted at six pharmacies using Kardia Mobile and Hartmann Veroval 2 in 1. A single-lead ECG was acquired by the placement of fingers on the pads. A cardiologist evaluated findings from both devices. Atrial fibrillation was identified in 3.60% and previously unknown AF was detected in 1.92% of the study participants. Sensitivity and specificity of the Kardia application in detecting AF were 66.7% (95% confidence interval [CI] 38.4-88.2%) and 98.5% (95% CI 96.7-99.5%), and for Veroval 10.0% (95% CI 0.23-44.5%) and 94.96% (95% CI 92.15-96.98%), accordingly. Inter-rater agreement was k = 0.088 (95% CI 1.59-16.1%). Mobile devices can detect AF, but each finding must be verified by a professional. The Kardia application appeared to be more user-friendly than Veroval. Cardiovascular screening using mobile devices is feasible at pharmacies. Hence it might be considered for routine use.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call