Abstract
In this phantom study, we tried to compare matching dosimetry differences between half-blocking of Co-60 and asymmetric collimation of the 4 MV photons during craniospinal irradiation. The dose distributions are compared and discussed. Firstly, some gaps with different sizes are left between cranial and spinal field borders. Secondly, the fields are overlapped in the same sizes. We irradiate the films located in water-equivalent solid phantoms with Co-60 and 4 MV photon beams. This study indicates that the field placement errors in ± 1 mm are acceptable for both Co-60 and 4 MV photon energies during craniospinal irradiation with half-beam block technique. Within these limits the dose variations are specified in ± 5%. However, the setup errors that are more than 1 mm are unacceptable for both asymmetric collimation of 4 MV photon and half-blocking of Co-60.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.