Abstract

A two-temperature humidity generator has been developed at the Centre for Metrology and Accreditation (CMA) in order to have a reliable humidity standard. The system has been tested and the analysis of uncertainty has been carried out. The disturbance of the saturation equilibrium, the accuracy of temperature measurements, pressure differences, leaks, desorption and contamination have been investigated. The measurement range of the system is from −40 to +60°C with an uncertainty of less than ±0.25°C ( k=2). The system was found suitable to be used as the CMA humidity standard.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.