Abstract

This study aimed to analyze the factors that impact the cloud point extraction of ciprofloxacin, levofloxacin, and moxifloxacin. The following independent variables were analyzed: Triton X-114 concentration, NaCl concentration, pH, and incubation temperature. The dependent variable studied was recovery. A central composite design model was used. The applied quantitation method was HPLC. The method was validated for linearity, precision, and accuracy. The results underwent ANOVA® analysis. The polynomial equations were generated for each analyte. The response surface methodology graphs visualized them. The analysis showed that the factor most affecting the recovery of levofloxacin is the concentration of Triton X-114, while the recovery of ciprofloxacin and moxifloxacin is most affected by pH value. However, the concentration of Triton X-114 also plays an important role. The optimization resulted in the following recoveries: for ciprofloxacin, 60%; for levofloxacin, 75%; and for moxifloxacin, 84%, which are identical to those estimated with regression equations-59%, 74% and 81% for ciprofloxacin, levofloxacin, and moxifloxacin, respectively. The research confirms the validity of using the model to analyze factors affecting the recovery of the analyzed compounds. The model allows for a thorough analysis of variables and their optimization.