Abstract

In this paper we have developed and evaluated a spherical phased array ultrasound applicator for deep tissue ablation. The 90-element prototype array has a 21cm aperture and an 18cm radius of curvature with a 5cm wide central hole. Annular distribution with circular elements is used to reduce the number of elements. The array is constructed with piezoelectric (PZT-8) circular planar elements that are 1.4cm in diameter and 0.2-cm thick. With the water-muscle propagation path, the array offers an effective beam focusing depth of at least 8cm in the muscle layer. Simulation results show that the array provides good beam focusing and steering capability over a cylindrical volume of approximately π×1×1×4 cm3 (up to 10mm off center over ranges from 15cm to 19cm). We also present its beam focusing and steering capability in deep tissue through a series of ex vivo experiments by measuring discoloration areas after sonications. The ex vivo experiments show a similar focal range as that found in the simulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.