Abstract

A graphite rod electrothermal vaporizer used to introduce microliter‐sized samples into a direct current plasma (DCP) atomic emission spectrometer is reported in this work. Several important experimental conditions were found to be important in achieving good analytical performance from the electrothermal vaporization (ETV‐DCP) system. A combination of lowered plasma electrode sleeve gas flow rates, when compared to that commonly used in the DC plasma jet, and relatively low carrier gas flow rates resulted in very good analytical performance. Relative precision for the ETV‐DCP instrument using B, Cd, Cu, Fe, and Pb solutions ranged between 4% and 10%. Limits of detection (LOD) lower than 100 pg were achieved for these elements, roughly an order of magnitude better than other ETV‐DCP studies that used commercial graphite boat/furnace combinations. In addition, good calibration linearity was observed, with 2–4 orders of magnitude linearity for the elements investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.