Abstract
The characteristics of the Ricketts maxillary canine retraction spring (Ricketts, 1974) have been derived using the complementary (strain) energy method. The theoretical predictions of the analysis have been confirmed by experiments on large planar models in a jig fitted with strain gauges capable of measuring the forces and couples developed when the springs were activated in a way simulating clinical use. The verified analysis was used to compare the horizontal stiffness and anti-tilt couple to retraction force ratios of the Ricketts with other simpler retraction components. The behaviour of the Ricketts mandibular sectional, as predicted by the analysis was also examined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.