Abstract

The use of Complex and High-risk Coronary Interventions (CHIPs) has increased in recent years. Both rotational atherectomy (RA) and hemodynamic support are important parts of CHIPs. This study aimed to retrospectively investigate the procedure results and clinical outcomes of intra-aortic balloon pump (IABP)-assisted RA in the contemporary drug-eluting stent era. All consecutive patients who received RA under in-procedure IABP assistance from April 2010 to March 2018 were analyzed retrospectively. A total of 63 patients (77.7 ± 10.1 years, 69.8% male) were recruited, of whom 51 underwent RA with primary IABP assistance and 12 underwent bailout IABP. RA could be completed in 61 (96.8%) of the patients. Overall, vessel perforation, profound in-procedure shock, and ventricular arrhythmia occurred in 1.6%, 4.8% and 3.2% of the patients, respectively. The in-hospital, 30-day and 90-day major adverse cardiac event (MACE) rates were 22.2%, 27.4% and 36.1%, respectively, mostly driven by mortality. The MACE rates were significantly higher in the bail-out group in the hospital (50.0% vs. 15.7%, p = 0.018) at 30 days (58.3% vs. 20.0%, p = 0.013) and 90 days (66.7% vs. 28.6%, p = 0.020). Bail-out IABP was associated with increased MACEs, implying that the use of IABP should be implemented at the beginning of RA if a complex procedure is anticipated.