Abstract

A theoretical study was performed to determine the accuracy and repeatability of multiple one-dimensional pulse sequences in the quantification of glutamine concentration at 3 T. Variable repeatability (12% to > 50%) and significant absolute error (-50% to +70%) were noted for the eight pulse sequences considered. Data acquired in vivo using three of the pulse sequences used for simulation matched the predicted repeatability well; among the pulse sequences considered, point-resolved spectroscopy (TE = 80 ms) offered minimal error and acceptable repeatability (12%) for brain glutamine measurements. Following correction for the expected bias of each pulse sequence, consistent glutamine measurements, in the 1-mM range, were reported with the three sequences. An explanation for the mismatch between in vivo (1)H MRS and in vitro (13)C/(1)H MRS at high field was attempted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.