Abstract

A high-purity germanium (HPGe) gamma-ray spectrometer and an Alpha-GUARD radon monitor are the activity and activity concentration transfer standards of 222Rn, respectively. The gaseous 222Rn standard source traceable to the NIM's absolute standardization apparatus of 222Rn was developed for calibrating these transfer standards. The calibration results of HPGe detector at three different distances show that the uncertainties of experimental efficiency are less than 0.77% (k = 1). Meanwhile, the Monte Carlo (MC) simulation was carried out at wider distances range for the HPGe calibration as well, and the deviation at same three distances between experiments and simulations results is within ±3.0%. The advantage of calibration using MC simulation and the possible reasons for the deviation were discussed in detail in this article. In addition, a reference activity concentration environment based on the gaseous 222Rn standard source was also developed to calibrate the radon monitor, and the typical uncertainty for calibration factor of Alpha-GUARD radon monitor could be reduced to 1.9% (k = 1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.