Abstract

A calibration system is described which is based on the use of a few precisely determined experimental values obtained from freezing-point baths. Characteristics of the individual thermocouples at intermediate points are obtained by passing empirical equations of prescribed form through the test values. A program is reviewed, by which a high-speed digital computer accomplishes the necessary conversions, curve fittings, comparisons of individual characteristics with arbitrary reference tables, and the printing out of a table of differences. Test results for a series of iron-constantan thermocouples, over the temperature range 32–1225 F, are presented to illustrate the use of the system and the uncertainties involved. Comparisons are drawn between these results and those obtained by other methods.