Abstract

An accelerometer-array calibration method is proposed in this paper by which we estimate not only the accelerometer offsets and scale factors, but also their sensitive directions and positions on a rigid body. These latter parameters are computed from the classical equations that describe the kinematics of rigid bodies, and by measuring the accelerometer-array displacements using a magnetic sensor. Unlike calibration schemes that were reported before, the one proposed here guarantees that the estimated accelerometer-array parameters are globally optimum in the least-squares sense. The calibration procedure is tested on OCTA, a rigid body equipped with six biaxial accelerometers. It is demonstrated that the new method significantly reduces the errors when computing the angular velocity of a rigid body from the accelerometer measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.