Abstract

1. An EEL mains-operated colorimeter has been calibrated to obtain rapid estimates of the concentration of spermatozoa in bull semen.2. A general linear calibration relationship has been calculated givingh, the estimated concentration of spermatozoa per ml. of the original semen, aswheredis the dilution factor andPdis the absorptiometer reading.3. Under the circumstances of the experiments described, the accuracy of estimates from the above relation was about 1%, and it is considered that the instrument may be effectively used in semen determinations for artificial insemination and in most cases where a rapid determination is required.4. Attention is drawn to the desirability of using standards and occasional check calibrations, and to the possibility that in routine use results may be slightly less accurate than those given above, which were obtained with laboratory determinations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.