Abstract

The neutron capture cross sections are very important in the field of nuclear device design and basic physics research. Hydrogen-free liquid scintillator such as C6D6 detectors are widely used in the neutron capture cross-sectional measurements for the low neutron sensitivity and fast time response. The Back-n white neutron source at China Spallation Neutron Source is the first spallation white neutron source in China, and it is suitable for neutron capture cross-sectional measurement. A C6D6 detector system was built in the Back-n experimental station. The pulse height weighting technique was used to determine the system’s detection efficiency. The response to gamma rays of the C6D6 detector was measured, and the energy resolution function was determined. Monte Carlo simulation with Geant4 code was carried out to get the weighting function of this C6D6 detector system. Additionally, the systematic uncertainty of the weighting function was also determined. According to the experimental and simulation results, this C6D6 detector system can be used to measure neutron capture cross section.