Abstract
During pain control by continuous epidural analgesia, the breakage of epidural catheter is a very rare complication. We experienced breakage of epidural catheter in patient of cancer pain during continuous epidural analgesia. We performed epidural catheterization between T12 and L1 level for pain control. After 6 weeks of epidural catheterization, abruptly patient complained severe pain. When we applied negative pressure to catheter by syringe before epidural injection of bolus dose, a few airs were aspirated. We strongly suspected breakage of catheter, and the breakage in 16-17 cm from epidural catheter tip was confirmed by normal saline administration via epidural catheter. After new epidural catheter inserted again, patient's pain was controlled by VAS 2-4 and another complications were not appeared. We must carefully manage the catheter because of complications of epidural catheter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.