Abstract

This work presents progress on development of the “Bladder pill”, a pressure monitoring implant for urological diagnosis. The device, measuring as little as 5x30mm, uses resonant inductive coupling for wireless powering and communication via an omnidirectional “3-D” receiving coil. The transmitting coil along with the battery-powered control and data logging circuitry is integrated into an external belt suitable for in-vivo testing in mini-pigs. All circuitry is realized using low cost off-the-shelf components, including the pressure sensor which is decapsulated to allow potting in PDMS. Powering and communication is successfully achieved at a distance up to 12cm orthogonally from the belt with data rates exceeding 10kbps and pressure sampling rate of 10Hz with < 2.5W power input. Finally, the results of a preliminary test involving a measurement of bladder contractions and voiding in anaesthised mini-pigs are presented and discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.