Abstract

The BiVACOR is a novel rotary Biventricular Assist Device (BiVAD) undergoing development to treat global end-stage heart failure. The design includes left and right vanes positioned on a shared rotating hub to form a double-sided magnetically and hydrodynamically suspended centrifugal impeller. The performance of the device was assessed in a pulsatile mock circulation loop replicating end-stage biventricular heart failure, and was shown to restore flow from pathological (2 L/min) to normal levels (5 L/min). A novel technique to balance the left/right outflow of the BiVAD was also investigated, for which a maximum relative left to right outflow differential of 1.8 L/min was achieved at normal physiologic afterloads. The in vitro results encourage device progression to in vivo animal studies. Successful development of this BiVAD will provide a suitably miniature device for patients who require biventricular assistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.