Abstract
A Bird Mk. 8 respirator has been modified to permit the use of PEEP, CPAP and IMV. Gas supply is from an oxygen blender to allow easy adjustment of the inhaled oxygen concentration. A special fitting was made to mount a flow meter which supplies the IMV reservoir. PEEP is generated by the Mk. 8 negative pressure generator mounted in opposition to exhalation and its magnitude can be adjusted by the negative pressure control of the respirator. This modified machine is simple, reliable and relatively inexpensive.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.