Abstract
The development of BiPo detectors is dedicated to the measurement of extremely high radiopurity in 208Tl and 214Bi for the SuperNEMO double beta decay source foils. A modular prototype called BiPo‐1 with 0.8 m2 of sensitive surface area has been running in the Modane Underground Laboratory since February 2008. The goal of BiPo‐1 is to measure the different components of the background and in particular the surface radiopurity of the plastic scintillators that make up the detector. The first phase of data collection has been dedicated to the measurement of the radiopurity in 208Tl. After more than one year of background measurement a surface activity of the scintillators of A(208Tl) = 1.5 μBq/m2 is reported here.Given this level of background, a larger BiPo3 detector having 3.25 m2 of active surface area, will able to qualify the radiopurity of the SuperNEMO selenium double beta decay foils with the required sensitivity of A(208Tl)<3–4 μBq/kg (90% C.L.) with a six month measurement. This detector is actually under construction and will be installed in the Canfranc Underground Laboratory mid 2011.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.