Abstract

After several years of development and continuous improvement of the BIPM Kibble balance, mass measurement of a 1 kg Pt/Ir artefact in terms of the Planck constant has been carried out. The paper describes the apparatus and its alignment, and explains the measurement and data processing. The balance operates in a one-mode and two measurement phase scheme using a bifilar coil. The mass value of the artefact was measured in vacuum with a relative standard uncertainty of 49 µg. This value differs from the value deduced from the calibration following the dissemination process after the redefinition of the kilogram by 15 µg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.