Abstract

Background: Injectable cardiac monitors (ICMs) are leadless subcutaneous devices for long-term monitoring of arrhythmias. The BIOTRONIK BIOMONITOR III is a novel ICM with a miniaturized profile, long sensing vector, and simplified implantation technique. Methods: R-wave amplitude was recorded immediately after implantation, the day after implantation, and after 3 months. Follow-up was scheduled after 3 months or after an event. All data from the ICM were retrieved. The anatomical position of the ICM was determined post-implantation and after 3 months. A patient questionnaire was conducted after 3 months. Results: In 36 patients (mean age 67 ± 13 years; 40% male) an ICM was inserted. Six patients were not included in the final analysis. The median time from skin cut to wound closure was 6 [IQR 5–7] minutes. Mean R-wave amplitude increased over time (0.73 ± 32 mV vs. 0.78 ± 0.38 mV vs. 0.81 ± 0.39 mV; p = ns). Three months after implantation, the ICM was in an anatomically stable position. In 14 (47%) patients, true episodes were detected. False arrhythmia alerts were detected in 13 (43%) patients. The total number of false detections was low, and the patient satisfaction rate was high. Conclusion: Implantation of the novel BIOMONITOR III is fast and uncomplicated; its sensing characteristics are excellent and improve over time, and patient satisfaction is high.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.