Abstract
Abstract An automated procedure is presented for the enzymatic determination of serum uric acid on both the AutoAnalyzer and the Robot Chemist. The procedure measures as the neocuproine complex, the difference in the amount of Cu+ formed by reaction of a Cu++-alkanolamine buffered solution with serum uric acid under precisely controlled conditions before and after uricase treatment of the serum. The difference is proportional to the true serum uric acid content. The elements contributing to the enzymatic reaction, the colorimetric reaction, and the elimination of interferences were investigated. Comparison of serum uric acid values obtained by this method with those obtained by ultraviolet spectrophotometry show very good agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.