Abstract
Transferrin and hemoglobin have been reported to oxidize L-ascorbic acid (AA) in vitro. The aim of this study was to determine whether or not physiological concentrations of serum transferrin (reference range 22-45 micromol/L) and hemoglobin (reference range 0-3.0 micromol/L) interfer with the measurement of AA in the serum. Transferrin (33 to 41 micromol/L) and hemoglobin (1.9 micromol/L) added to freshly pooled serum significantly decreased measured AA in the serum (p < 0.05). However, we found that the magnitude of the decrease in AA due to transferrin at concentrations within the reference range or up to 80 micromol/L was inconsequential, and had no clinical importance in diagnosing a low AA concentration. Hemoglobin at concentrations within the reference range had little affect on the serum AA measurement. However, when serum specimens were stored at 4 degrees C for more than 1.5 h, the magnitude of the decrease in AA due to hemoglobin at physiological concentrations may cause a misleading clinical diagnostic evaluation of low AA concentration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.