Abstract

A sensitive, direct spectrofluorimetric method for the trace determination of terbium with use of trimesic acid (TMA) has been developed. The reaction conditions for the fluorescence system of terbium with TMA were studied. The terbium ion can form a stable binary chelate with TMA, having a ratio of 1:1 in the pH range 3.5–6.5. The maximum excitation and emission wavelengths are 260 nm and 545 nm for the terbium chelate, respectively. The reaction is instantaneous and the fluorescence intensity of the terbium chelate remains stable from 0.25 to 4 h. Under the optimal experimental conditions the fluorescence intensity is a linear function of concentration in the range 0.0248–6.35 μg ml −1 of terbium. The relative standard deviation is still within ±4% in the presence of one-thousandfold amounts of the other lanthanide ions, and common foreign ions hardly interfere in the determination. The method can be employed for the determination of trace amounts of terbium in rare earth ores or oxides because of its high sensitivity and selectivity with good reproducibility and accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.