Abstract
A uricase-peroxidase coupled system, for the determination of uric acid, applied to a CentrifiChem-500 centrifugal fast analyser is described. Relatively large amounts of ascorbic acid, due to the inclusion of an ascorbate oxidase urine diluent, and hemoglobin appear not to interfere with the procedure while the incorporation of potassium ferrocyanide into the reagents has led to the near-total elimination of bilirubin interference. The incubation period is relatively short compared with other similar procedures and the one reagent system has made the procedure simple to perform. The use of sodium 2-hydroxy-3,5-dichlorobenzenesulfonate and 4-aminoantipyrene in the oxidative coupling reaction has incorporated the advantages of increased sensitivity, over phenol-4-aminoantipyrene systems, as well as the amenability of the reagent towards lyophilization or “dry-fill”.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.