Abstract
The Ar spectral lines are suppressed in glow discharge source atomic emission spectroscopy by a double voltage modulation technique with a supplementary electrode. The voltage is modulated between two levels, typically 350 and 700 V. At the lower voltage level mainly the Ar emission occurs where at the higher voltage level both the Ar and the metal atoms sputtered from the sample contribute to the emission. The power supply of the supplementary electrode is switched on when the glow discharge passes from operation at 700 V to 350 V. The intensity of argon gase lines is regulated by the current applied to the supplementary electrode at that period. Therefore, there is no need for electronical amplification and lower voltage adjustment of two operation modes. This modified DVM technique has been applied to the determination of Si and Cu in Al- samples. The suppression of Ar lines is possible, and the elements can be determined without Ar interferences.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.