Abstract

Purpose: Maturation of an arterio-venous fistula (AVF) frequently fails, with low postoperative fistula flow as a prognostic marker for this event. As pulsed wave Doppler (PWD) is commonly used to assess volume flow, we studied the accuracy of this measurement in the setting of a radiocephalic AVF. Methods: As in-vivo validation of fistula flow measurements is cumbersome, we performed simulations, integrating computational fluid dynamics with an ultrasound (US) simulator. Flow in the arm was calculated, based on a patient-specific model of the arm vasculature pre and post AVF creation. Next, raw ultrasound signals were simulated, from which the Doppler spectra were calculated in both a proximal (brachial) and a distal (radial) location. Results:The velocity component in the direction of the US beam, in a centred, small, sample volume, can be captured accurately using PWD spectrum mean-tracking. However, deriving flow rate from these measurements is prone to errors: (i) the angle-correction which is influenced by the radial velocity components in the complex flow field; (ii) the largest error is introduced due to a lack of knowledge on the spatial flow profile.