Abstract

Children with cyanotic congenital heart disease (CCHD) live with oxyhemoglobin saturations that are typically expressed as percentages in the range of 70s and 80s. Peripheral pulse oximetry (measurement of SpO2) performs poorly in this range and yet is widely used to inform clinical decisions in these patients. The reference standard is co-oximetry of arterial samples (SaO2). In this study, 515 paired measurements of SpO2 and SaO2 were taken from 19 children who had undergone palliative cardiac surgery. SpO2 (Masimo SET LNCS Neo pulse oximeter) overestimated oxyhemoglobin saturation in 82% of measurements (mean 4.6% ± 6.6%). There was a strong negative correlation between mean bias and SaO2 ( r = -.96, P = .002, 95% confidence interval: -0.99 to -0.68). The results raise a concern that critical hypoxemia may go undetected and untreated if pulse oximetry is relied upon as the primary means of assessing oxyhemoglobin saturation in children with CCHD. Strong preference must be given to co-oximetry of arterial samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.