Abstract
We tested whether operators using manometers attached to self-inflating bags could accurately deliver set targeted peak inspiratory pressures (PIPs) compared to the Neopuff(™) T-piece resuscitator (TPR). Participants provided positive pressure ventilation to a leak-free neonatal test lung at a rate of 60 inflations/min and a flow of 8L/min. Participants used three manometers attached to self-inflating bags and a Neopuff(™) TPR to target PIPs of 20, 30 and 40cmH2 O on each device. Mean PIPs delivered with each manometer were compared to the 'gold standard' Neopuff(™) TPR. In total, 13991 inflations delivered by 20 participants were analysed. At all target PIPs, the mean PIP delivered using the Mercury Medical manometer attached to a Laerdal self-inflating bag was significantly higher by 5cmH2 O (p<0.01) than the Neopuff(™) TPR. The PIP delivered using both the Ambu(™) and Parker Healthcare manometers attached to their respective devices was similar to that delivered by the Neopuff(™) TPR at all targeted PIPs. Accurately targeted PIPs can be achieved when a manometer specifically designed for use on a self-inflating bag is used during manual ventilation. This may be useful in settings where access to a Neopuff(™) TPR or a gas flow source is limited.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.