Abstract
To assess the accuracy of continuous glucose monitoring (CGM) in medical intensive care unit (MICU) patients. A Medtronic Enlite® sensor accuracy was assessed versus capillary blood glucose (CBG) and plasma glucose (PG) using the mean absolute relative difference (MARD), surveillance error grid (SEG) analysis and modified Bland-Altman plots. Using CBG as a reference, MARD was 6.6%. Overall, 99.7% of the CGM readings were within the "no risk" zone. No significant differences in accuracy were seen within vasopressor subgroups. Using PG as the reference, MARD was 8.8%. The surveillance error grid analysis showed 95.2% of glucose readings were within the "no risk" zone. There were no device-related adverse events. The CGM sensor showed acceptable accuracy in MICU patients, regardless of vasopressor use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.