Abstract

Visual blood loss estimation often underestimates blood loss. In this study we sought to determine the effect of calibrated drape markings on blood loss estimation in a simulated vaginal delivery. Subjects were randomized to estimate simulated blood loss (300, 500, 1000, and 2000 mL) in calibrated or noncalibrated vaginal delivery drapes and then crossover. Visual blood loss estimation with noncalibrated drapes underestimated blood loss, with worsening accuracy at larger volumes (16% error at 300 mL to 41% at 2000 mL). The calibrated drape error was <15% at all volumes. Calibrated vaginal delivery drapes improve blood loss estimation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.