Abstract

The 6.9 F semirigid ureteroscope has a hybrid design that shares the rigid ureteroscope body with the fiberoptic imaging bundle of the flexible ureteroscopes. This allows more space for the inner channel, so that there are two working channels of 3.4 F and 2.3 F. This ureteroscope was used in 65 cases on 57 patients without dilatation of the intramural ureter. It was mainly used for the management of ureteral calculi (29 cases) and for visual diagnostic purposes (28 cases). The double-channel design proved extremely helpful in lithotripsy, while the ability to access the ureter without dilatation gave a consistently clearer picture for diagnostic purposes. Forty-two cases were done under local anesthesia with sedation. In I patient a minor perforation developed secondary to use of electrohydraulic lithotripsy (EHL). Postoperatively, 7 patients complained of colic, and in 5 patients a low grade fever developed, while 2 patients had both.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.