Abstract

The technical innovation of boron-doped diamond on top of a silicon layer as alpha nuclide measurement device offers promising advances, since the energy-resolved measurement takes place in-situ in water without matrix separation. Several experiments were performed to detect disturbance factors by investigating the yield of 241Am in the presence of main or minor drinking water elements. The results show that an activity was detected by the sensor in any case. Generally, the yields were higher with NaNO3 as electrolyte than with Na2SO4, due to speciation and higher hydrogen formation. A reduced yield was observed for nitrate as electrolyte and maximum concentrations of Ca, Mg, PO4, and F in drinking water. The generally high standard deviation of the measurement, which is mainly due to hydrogen evolution, mass attenuation, and surface site occupation, does not allow for an exact determination of a present α-contamination. However, a significant aqueous alpha activity can be detected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.