Abstract

Modern ceramic bearings used in total hip arthroplasty (THA) consist of a femoral head (ball) articulating inside a hemispherical acetabular cup (socket); the ball and socket are made of alumina (Al(2)O(3)) and Al(2)O(3)-based composite materials. In the present study, total hip bearings were made from a different ceramic material, silicon nitride (Si(3)N(4)), by sintering and hot isostatic pressing of powders. The resulting material had improved mechanical properties over modern Al(2)O(3) THA bearings, with a flexural strength of 920 +/- 70 MPa, a Weibull modulus of 19, and a fracture toughness of 10 +/- 1 MPa m(1/2). Unlike zirconia-based ceramics that have also been used in THA, accelerated aging of Si(3)N(4) did not adversely affect the flexural strength. In simulated wear tests, Si(3)N(4) acetabular cups produced low-volumetric wear whether articulating against Si(3)N(4) or cobalt-chromium (CoCr) femoral heads. The results of this investigation suggest that Si(3)N(4) may allow improved THA bearings that combine the reliability of metal femoral heads with the low wear advantages of ceramic materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.