Abstract

The test apparatus of infrared emissivity used in powders industry was researched by controlling the temperature accurately and collecting the signals of surface temperature of two standard samples and a test sample. The tourmaline samples from ten mines had been tested by the new test apparatus. Compared with the results from FTIR, it was shown that average value of relative error was 0.378% and repeatability error was less than 0.6% after correcting by the coefficient kc = 0.958.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.