Abstract

A person-wearable dust monitor that provides nearly real-time, mass-based readings of respirable dust was developed for use in underground coal mines. This personal dust monitor (PDM) combined dust sampling instrumentation with a cap lamp (and battery) into one belt-wearable unit, with the air inlet mounted on the cap lamp. However, obsolescence of belt-carried cap lamp and batteries in coal mining ensued and led end users to request that the cap lamp and battery be removed from the PDM. Removal of these components necessitated the design of a new air inlet to be worn on the miner’s lapel. The revised inlet was tested for dust collection equivalency against the original cap-mounted inlet design. Using calculated inlet respirable fractions and measured dust mass collection, the performance of the two inlets is shown to be similar. The new inlet requires a 1.02 factor for converting dust masses obtained from it to equivalent masses collected from the original inlet.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.