Abstract

Vestibulo-ocular reflex rotational chair testing in the high-frequency range is seldom performed because it requires specialized and powerful systems. But today a new method of sweep-frequency vestibulo-ocular reflex testing, the Vestibular Autorotation Test system (Western Systems Research, Inc., Pasadena, Calif.), based on active head movements increasing from 2 to 6 Hz, is available on the market. The goal of this study was to evaluate the test-retest variability of this test in healthy subjects. Twelve young adults (22 to 42 years old) without any history of auditory or vestibular dysfunction were included in the study. Subjects underwent five tests under standardized conditions with a 1-week interval. Each test consisted of three measurements of the gain and phase of the vestibulo-ocular reflex in the horizontal and vertical planes. Statistical analysis shows that the test-retest reliability of the Vestibular Autorotation Test is poor. Therefore this method cannot be used routinely to evaluate precise vestibulo-ocular reflex anomalies. (Otolaryngol Head Neck Surg 1997;117:704-7.)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.