Abstract
The purpose of this study was to investigate the test-retest reliability of contact heat-evoked potentials (CHEPs) in neurologically healthy subjects from cervical dermatomes (C4-C8). Seventeen individuals underwent test-retest examination of cervical CHEPs. Peak latencies and peak-to-peak amplitude of N2-P2 and ratings of perceived intensity were analyzed using test-retest reliability statistics (intraclass correlation coefficients [ICCs] and Bland-Altman confidence parameters). For comparison, a group of similar age and gender was also examined with dermatomal somatosensory-evoked potentials (dSSEPs, n = 17). The ICC values for CHEP latency and amplitude parameters were significant (P < 0.05) and corresponded to at least "fair" reliability, while peak-to-peak amplitude demonstrated "substantial" (≥0.81) reliability for all dermatomes. The coefficients of repeatability (i.e., 2SD of the difference between examinations) confirm that CHEPs and dSSEPs are reliable according to measures of latency. Superior peak-to-peak amplitude test-retest reliability was found for CHEPs. In conclusion, the test-retest reliability of dSSEP and CHEP parameters supports the fact that these outcomes can be used to objectively track changes in spinal conduction in the dorsal column and spinothalamic tract, respectively. The reliable acquisition of CHEPs may depend on the intensity of the sensation reported by the subject for a given area of skin stimulated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.