Abstract

To compare the test-retest reproducibility of three variants of arterial spin labeling (ASL): pseudo-continuous (pCASL), pulsed (PASL) and continuous (CASL). Twelve healthy subjects were scanned on a 3.0T scanner with PASL, CASL, and pCASL. Scans were repeated within-session, after 1 hour, and after 1 week to assess reproducibility at different scan intervals. Comparison of within-subject coefficients of variation (wsCV) demonstrated high within-session reproducibility (ie, low wsCV) for CASL-based methods (gray matter [GM] wsCV for pCASL: 3.5% ± 0.02%, CASL: 4.1% ± 0.07%) compared to PASL (wsCV: 7.5% ± 0.06%), due to the higher signal-to-noise ratio (SNR) associated with continuous labeling, evident in the 20% gain in temporal SNR and 58% gain in raw SNR for pCASL relative to PASL. At the 1-week scan interval, comparable reproducibility between PASL (GM wsCV 9.2% ± 0.12%) and pCASL (GM wsCV 8.5% ± 0.14%) was observed, indicating the dominance of physiological fluctuations. Although all three approaches are capable of measuring cerebral blood flow within a few minutes of scanning, the high precision and SNR of pCASL, with its insensitivity to vessel geometry, make it an appealing method for future ASL application studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.