Abstract

This study aimed to evaluate the test–retest reliability of a new device for assessing ankle joint kinesthesia. This device could measure the passive motion threshold of four ankle joint movements, namely plantarflexion, dorsiflexion, inversion and eversion. A total of 21 healthy adults, including 13 males and 8 females, participated in the study. Each participant completed two sessions on two separate days with 1-week interval. The sessions were administered by the same experimenter in the same laboratory. At least 12 trials (three successful trials in each of the four directions) were performed in each session. The mean values in each direction were calculated and analysed. The ICC values of test–retest reliability ranged from 0.737 (dorsiflexion) to 0.935 (eversion), whereas the SEM values ranged from 0.21° (plantarflexion) to 0.52° (inversion). The Bland–Altman plots showed that the reliability of plantarflexion–dorsiflexion was better than that of inversion–eversion. The results evaluated the reliability of the new device as fair to excellent. The new device for assessing kinesthesia could be used to examine the ankle joint kinesthesia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.