Abstract

This study aimed to evaluate a recently developed equipment test method by assessing the safe and accurate functioning of the Abbott Optium FreeStyle H portable blood glucose monitor for use in the Alfred Hospital's hyperbaric chamber. The results of this study indicate that the test method can be used successfully to evaluate instruments and/or devices for use in the hyperbaric environment. The evaluation initially found that this particular glucose monitor contained a lithium battery which can be hazardous when used in the hyperbaric environment. However, upon further inspection it was determined the battery posed minimal risk for fire and explosion due to its small capacity and design application. The results indicate that the Abbott Optium FreeStyle H blood glucose monitor operated normally when used in the hyperbaric chamber. This glucometer was found to perform within the calibration specification requirements for accuracy at all stages of a typical hyperbaric treatment and as such the Abbott Optium FreeStyle H blood glucose monitor was deemed safe for use in the hyperbaric chamber at the Alfred Hospital.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.