Abstract

In its statement on tissue reactions approved on 21st April 2011, the International Commission on Radiological Protection (ICRP, 2012) reviewed its recommendation concerning the equivalent dose limit for the eye lens and reduced the dose limits for occupationally exposed persons to 20 mSv in a year, averaged over defined periods of 5 years, with no single year exceeding 50 mSv. This limit was approved and written down in the new EURATOM (European Atomic Energy Community) directive 2013/59 and in the IAEA (International Atomic Energy Agency) BSS (Basic Safety Standard) of July 2014.For that reason, the necessity to monitor the eye lens may become more important than it was before. However, specially dedicated dosemeters for the dose quantity Hp(3) are using very rarely. Commonly use are only whole body personal dosemeters for the personal dose equivalent quantities Hp(10) worn on the trunk and ring dosemeters worn on finger to measure the quantity Hp(0.07).Therefore, in this work it was investigated whether dosemeters from routine use calibrated in terms of Hp(10) and Hp(0.07) and worn on thyroid collar and protective apron could deliver similar results like dedicated eye lens dosemeter worn close to the eyes.The results show that the best method if dedicated eye lens dosimeters is not used is to measure doses in terms of Hp(0.07) on the thyroid collar (Pearson product, r=0.85). Obtained results shows also importance of proper localization of eye lens dosimeter (close to the eye, from side of the X-ray source).