Abstract

Abstract The radial strength of balloon expandable stents represents a key property for a successful recanalization of sclerotic blood vessels. This study focuses on the most commonly used method for investigation of radial strength and radial stiffness via segmented head test setup. A custom made user software was used for evaluation of the radial force curves considering requirements of international standards such as ASTM F3067. Contributing factors during measurement such as friction and test setup deformation as well as the single cycle and multi cycle approach were addressed and discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.