Abstract
The official requirements for sutures in Australia are those of British Pharmacopoeia (B.P.). The results of a survey conducted by this Laboratory indicate that the B.P. test and specifications for knot pull strength are no longer appropriate for sutures currently available in this country. It is suggested that tensile strength measurements on sutures should be carried out, without prior soaking, using the load cell type of constant rate of extension apparatus rather than the pendulum type tester specified in the B.P. Use of a simple knot is suitable for testing synthetic sutures, but the surgeon's knot is preferred for catgut. All products tested easily met both the B.P. and the United States Pharmacopeia (U.S.P.) requirements for tensile strength.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.