Various techniques and electrode catheters have been developed for effective and safe temporary transvenous cardiac pacing. In neonates and children, lead dislodgment and myocardial perforation have been attendant risks with passive fixation leads. We report our experience with the recently introduced Medtronic 6416 active fixation bipolar lead in a small group of children. Between January 2000 and February 2002, 4 children (2 boys, 2 girls) underwent temporary transvenous pacing with the Medtronic 6416 lead. The median age at implantation was 4.25 years (range: 5 days-13 years), median weight was 14.5 kg (range: 2.5 kg-54 kg) and the median duration of temporary pacing was 19.5 days (range: 5-38 days). The indication for temporary pacing was AV block (n = 2), sinus arrest (n = 1) and sinus node dysfunction (n = 1). The acute pacing threshold was <1.5 mA with no significant rise in the follow-up period. The venous access for lead placement was femoral vein (n = 2), internal jugular vein (n = 1), and subclavian vein (n = 1). There were no complications related to implantation or in the follow-up period. The Medtronic 6416 lead appears to be a safe and effective tool in temporary transvenous pacing in neonates and children.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.